Emerson 1500 Marine Instruments User Manual


 
OCT 2013 DANIEL™ SERIES 1500 TURBINE METER
SPECIFICATIONS2-4
Meter Performance
Linearity
± 0.15% (Linearity options of ± 0.1% and ± 0.07% available on certain sizes).
Repeatability
± 0.02% at any point throughout the extended minimum to extended maximum flow
range.
Pressure Drop
At maximum flow of normal flow range on water 4 psi.
Pressure Range
ANSI 150 - 1500 # depending upon size
Viscosity and Specific Gravity
Low specific gravities or high velocities will reduce flow range of meter. Consult Daniel
Measurement engineers.
Preamplifier Performance
Inputs
Supply voltage: 10-30 Vdc
Sensor Type: Reluctance
Type: Sine Wave
Amplitude: 40 mV peak to peak minimum
Outputs
Powered Pulse Output
Type: Square Wave
Frequency Range: 0 to 5 kHz
Amplitude: 0 to 5 V
1000 Ohm internal pull-ups, 20 mA, max.
Variable Voltage Output
Type: Square Wave
Frequency Range: 0 to 5 kHz
Amplitude: 0 to Supply Voltage
1000 Ohm internal pull-ups, 70 mA, max.
Open Collector Output (Per Channel)