Invacare 1110532 SONAR User Manual


 
SECTION 9—DIAGNOSTIC CODES
MK
5
™NX™Electronics 34 Part No 1110532
SECTION 9—DIAGNOSTIC CODES
What Are Diagnostics Codes?
The joystick information gauge and the Remote Programmer give indications of the type
of fault or error detected by the control module. When a fault is detected, the wheelchair
will stop and not drive. All of the lights on the information gauge will begin to flash (SPJ
and CSPJ joysticks) or the service indicator light will flash (SPJ+ and SPJ+ w/ACC
joystick). The number of flashes indicates the nature of an abnormal condition. An error
code and a quick description of the fault will begin to scroll across the Remote
Programmer display. If multiple faults are found, only the first fault encountered by the
control module program will be displayed. Refer to the Power Wheelchair Service
Manual for detailed troubleshooting and repair instructions. A table of the diagnostics
codes and their causes follows.
*NOTE: The fault log displays a four digit number. The first two digits are the diagnostic code and
the remaining two digits are the sub code.
NUMBER OF FLASHES
DIAGNOSTICS CODE
ERROR CODE
DESCRIPTION
SUB CODE*
DETAILS OF
ERROR CODE
POSSIBLE
SOLUTION
1 E 01 User Fault 00 Stall Timeout or user error. Release joystick to neutral and try again.
2 E02 Battery Fault 00 Recharge batteries or replace. Check the batteries and cable. Try charg-
ing the batteries. Batteries may require
replacing.
3 E03 Left Motor Fault 00 Left Motor Short Circuit Check the left motor, connections and
motor cable.
01 Left Motor Open Circuit
02 Left Motor Connection Fault
B-
03 Motor Terminal Connected
to B+
04 Left Motor Voltage Fault
05 Left Motor Bridge Fault
06 Too Many Hardware Current
Limit Events
07 Current Offset Out of Range
08 Hardware Current Limit Fault