Konica Minolta CS-2000 Marine Radio User Manual


 
82
Explanation
Should any errors be found in this instrument, try the corrective actions shown in the
following table. If this does not help, it’s possible the instrument is broken. Please
contact the nearest KONICA MINOLTA authorized service facility with the error
number and version of your instrument. To confirm the instrument version, see p.23.
Error
No.
Symptom Item to Check Corrective Action
Reference
Page
1 No display on the LCD
after power is turned
on.
Has the AC adapter been properly
plugged into the AC outlet?
Connect the AC adapter.
21
Has the AC adapter been
connected to this instrument?
Connect the AC adapter.
21
Has a wrong AC adapter
been connected?
Be sure to use the AC adapter and
power cord supplied as a standard
or optional accessory (AC-A312).
21
Is AC power source
within rated scale?
Use within ±10% of the nominal
voltage.
21
2 Nothing is visible
through the finder.
Is the lens cap still attached
to the objective lens?
Remove the lens cap.
8
Is the ND filter attached
to the objective lens?
Use the ND filter when the
luminance of that being
measured is too high.
9, 45
Is the ND eyepiece filter
attached to the finder?
Use the ND eyepiece filter when
luminance of that being
measured is too high.
9, 45
3 Nothing is displayed on
the LCD.
Is the backlight set to
OFF?
Press
BACKLIGHT
key to turn
ON the backlight.
14, 49
Has the backlight been
set to OFF during
measurement?
In the menu operation, set the
backlight to ON during
measurement
49
4 Does not accept key
operation.
Has the remote mode
been set?
Press
ESC
key to cancel the
remote mode.
73
Have you pressed a
disabled key?
Press the correct key.
5
Measurement unavailable
even if the measurement
button is pressed.
Has a screen other than
MENU been displayed?
Conduct measurement when the
MEAS screen is displayed. 53
6 The entered value for
target color is different
from the one that will be
displayed after setting.
A 1-digit difference may be
found due to calculation error.
61
7 Measurement values
do not appear.
Is there data? Conduct measurement. 52
Has the color space
mode become color
temperature?
Color temperature is displayed
as "_ _ _" if it is far from the
blackbody locus. Change the
display using the other color
space mode and check.
39
Did you interrupt
measurement?
Conduct measurement again.
52
Error Check