Konica Minolta CS-2000A Marine Radio User Manual


 
79
Explanation
Use of the data management software CS-S10w Professional in the standard accessories
enables simple measurement of the color of the object. Measured data is evaluated based
on luminance which has been stored as light source data in CS-S10w. See instruction
manual of CS-S10w for details.
Set while calibration plate (optional accessory) and object on the same position with
the same angle from this instrument. Uniform illumination and measurement conditions
of white calibration plate and those of object. If not, measurement data may vary,
causing incorrect data.
Keep illumination light source as stable as possible with fixed voltage power source
during measurement.
Measurement of Object Color
Measurement Procedure
Necessary Setting for Object Color Measurement
1. Set one or more tungsten lumps
or equivalent as illumination
source toward white calibration
plate as in the right illustration.
Set this instrument vertical to white
calibration plate.
Keep the angle between illumination
light source and white calibration plate
at 45°.
2. Start up this instrument, PC, and
software CS-S10w.
White Calibration
3. Set the document type to “object color” with CS-S10w.
4. Conduct white calibration with CS-S10w.
Measurement of Object
5. Set object on the same position and at the same angle as white
calibration plate.
6. Conduct measurement with CS-S10w.
See instruction manual of data management software CS-S10w for details.
White Calibration
Plate
Illumination Light Source
Illumination Light Source
45°
45°