Konica Minolta CS-2000A Marine Radio User Manual


 
81
Explanation
Error Message
Error message appears on the LCD when this instrument does not operate normally. Below
table shows type of error message, its description and corrective action respectively.
Error Message Cause (Description) Corrective action
1 MEMORY ERROR
Data stored in ROM is
damaged.
Turn on the power switch again.
If the symptoms are not remedied,
please contact the nearest
KONICA MINOLTA
authorized service facility.
2 OVER
Luminance of measuring
object is higher than the
available range.
Flicker of measuring object
is high.
Use the ND filter and re-measure.
Decrease the measuring area and
re-measure.
If the symptoms are not remedied,
please contact the nearest
KONICA MINOLTA authorized
service facility.
Set the flicker cycle in INT SYNC
mode, or input the flicker signal in
EXT SYNC mode.
3 SYNC ERROR
Input signal in EXT SYNC
mode is not at CMOS level.
Input signal in EXT SYNC
mode exceeds 200 Hz.
Input signal in EXT SYNC
mode is less than 20 Hz.
Input a signal with 3.5 to 5.0 V at
high level; 0.0 to 1.5 V at low level.
Set the frequency value divided by
an integer in INT SYNC mode and
re-measure.
Set the integral multiple value of
the input signal cycle in INT SYNC
mode and re-measure using the
MULTI INTEG-NORMAL mode or
MULTI INTEG-FAST mode.
Set the integral multiple value of the input
signal cycle in INT SYNC mode and re-
measure using the MANUAL mode.
4 VIEWING-ANGLE
SELECTOR ERROR
Measurement was
performed when the
measuring angle selector
was in the wrong position,
or its position was changed
during measurement.
Switch the measuring angle
selector and re-measure. Or, do
not operate the measuring angle
selector during measurement.
If the symptoms are not remedied,
please contact the nearest
KONICA MINOLTA authorized
service facility.
5 TEMPERATURE
ERROR
The ambient temperature for the
measurement device is too high,
and an internal temperature of
the sensor becomes abnormal.
Decrease the ambient temperature
to enable cooling to within the
specified temperature.
6 FAN ERROR
The cooling fan stops, or the
cooling device has an error.
Decrease the ambient temperature
to enable cooling to within the
specified temperature.
Turn off the power switch once,
turn it on again after a while.
If the symptoms are not remedied, the
fan must be repaired. Please contact
the nearest KONICA MINOLTA
authorized service facility.