Emerson Process Management ETC01035 Marine Instruments User Manual


 
MLT 2 / BINOS
®
100 F
ATEX Instruction Manual
ETC01035
January 2004
Emerson Process Management Manufacturing GmbH & Co. OHG1 - 6
1-3-4 Purge Gas Conditions
1-3-4 Purge Gas Conditions
Purge gas: Inert gas (e.g. Nitrogen)
Air (from an ex-free zone)
Temperature: As ambient, but min. 20 °C to 35 °C.
Input pressure at
inlet of EEx p device: 2.000 to 4.000 hPa (2 to 4 bar)
Medium has to be dry and free from
dust, oil, corrosive or aggressive
components!