2-7-1 Cable Gland Assembly Instruction
3. Feed cable through
gland nut and into
fixing element
4. Put the shielding net
over the element the
way that it covers the
o-ring 2 mm.
1. Strip the cable
insulation
2. Uncover the
shielding
5. Stick the fixing
element into the
neck and fix the
gland.
2-7-2 Signal Inputs and Outputs
Signal inputs and outputs by default are non-
intrinsically safe!
DIN rail installable converters (option) are
available to provide intrinsically safe signals
When the analyzer is out of order or if the pressurization unit shuts off all
non-intrinsically safe inputs and outputs connected to external equipment
MUST be shut off too!
This ensures that no hazardous voltages are present within the analyzer
enclosure when not pressurized.
EXPLOSION HAZARD !
(contact your local sales office for detailled
information).