Fluke 199 Marine Radio User Manual


 
Specifications
DMM Measurements on Meter Inputs
9
95
Diode
Maximum Voltage Reading....................................2.8 V
Open Circuit Voltage ..............................................<4 V
Accuracy.............................................
±(2 % +5 counts)
Measurement Current ............................ 0.5 mA,
±20 %
Temperature (TEMP)
With Optional Temperature Probe
Ranges (
°C or °F) ............................... -40.0 to +100.0 °
-100.0 to +250.0 °
-100.0 to +500.0 °
-100 to +1000 °
-100 to + 2500 °
Probe Sensitivity ..........................1 mV/°C and 1 mV/°F
DC Voltage (VDC)
Ranges....500.0 mV, 5.000 V, 50.00 V, 500.0 V, 1100 V
Full Scale Reading ..................................... 5000 counts
Accuracy..........................................
±(0.5 % +5 counts)
Normal Mode AC Rejection at 50 or 60 Hz
±1 % >60 dB
AC Voltage (VAC)
Ranges ....500.0 mV, 5.000 V, 50.00 V, 500.0 V, 1100 V
Full Scale Reading ..................................... 5000 counts
Accuracy
15 Hz to 60 Hz ...............................
±(1 % +10 counts)
60 Hz to 1 kHz.............................
±(2.5 % +15 counts)
For higher frequencies the frequency roll off of the
Meter input starts affecting accuracy.
Normal Mode DC Rejection................................>50 dB
AC+DC Voltage (True RMS)
Ranges ....500.0 mV, 5.000 V, 50.00 V, 500.0 V, 1100 V
Full Scale Reading ..................................... 5000 counts
Accuracy
DC to 60 Hz ...................................
±(1 % +10 counts)
60 Hz to 1 kHz.............................
±(2.5 % +15 counts)
For higher frequencies the frequency roll off of the
Meter input starts affecting accuracy.
All accuracies are valid if the waveform amplitude is
larger than 5 % of full scale.