Fluke 199 Marine Radio User Manual


 
Fluke 192/196/199
Users Manual
74
Changing the Auto Set Options
With the next procedure you can choose how auto set
behaves when you press the
AUTO (auto set) key.
1 Display the USER key labels.
2 Open the User Options menu.
3 Open the Auto Set Adjust
menu.
If the frequency range is set to > 15 Hz, the Connect-and-
View function responds more quickly. The response is
quicker because the test tool is instructed not to analyze
low frequency signal components. However, when you
measure frequencies lower than 15 Hz, the test tool must
be instructed to analyze low frequency components for
automatic triggering:
4 Select Signal > 1 Hz, then jump
to Coupling.
With the coupling option you can choose how auto set
behaves. When you press the
AUTO (auto set) key, the
coupling can either be set to dc or left unchanged:
5 Select Unchanged.
Note
The auto set option for the signal frequency is
similar to the automatic trigger option for the
signal frequency. (See Chapter 5: “Automatic
Trigger Options”). However, the auto set option
determines the behavior of the auto set function
and shows only effect when you press the auto
set key.