54 Micro Motion 7827 Digital Viscosity Meter
General Maintenance
Installation dimensions Check the key dimensions
as specified in Chapter 2
for your installation.
• Your key dimensions should match those
detailed in the Installation chapter for your
meter for your mounting configuration.
Product build up on the tines Product which has stuck
to the tines
• Remove with a suitable solvent.
Corrosion or erosion on the tines Scratches, pits or a
change in surface finish of
the tines.
• Recalibration is highly recommended.
• Where corrosion is a known problem, periodic
recalibration is recommended at time
intervals to suit the rate of corrosion of the
tine material.
Table 7-4 Installation checks continued
What to check: What to look for: What to check next: