Deltech Fitness 6 1 0 0 Marine Sanitation System User Manual


 
21
8 Cleaning and Inspection
Procedures
Use any of the following solutions to clean the
pump and accessories:
Soap solution
Benzalkonium chloride concentrate (0.13%)
Glutaral concentrate, USP (2%)
10 percent solution of household bleach (one
part household bleach to nine parts water)
Alcohol, USP (93%)
Isopropyl Alcohol, USP (99%)
PDI Super Sani-Cloth
®
Mada Medical MadaCide
1. Dampen a soft, lint-free cloth with cleaning
solution. Apply the solution to exterior
surface of the pump or accessory. Do not
allow the solution to soak into the pump or
accessory.
2. Wipe the entire surface dry with another
soft, lint-free cloth. Allow the pump to dry
completely before use.
Visual Inspection
Visually inspect the pump for any damage to
the LCD, occlusion sensor seals, valves and
expulsor, reservoir hinge area, latch, lock,
cassette sensors (3), keyboard, indicator
lights, power jack, data in/out jack, air
detector port cover or air detector, and
housing. If any damage is noted, the pump
should be returned for service.
Check the battery door for proper operation.
It should not be broken or damaged. The
mating tabs on the pump housing should not
be broken or damaged.
Examine the battery compartment for dam-
age. If the battery contacts appear corroded,
clean them with a cotton swab and isopropyl
alcohol. If the battery contacts appear to be
bent or pushed in, straightening may be
possible with a small screwdriver or other
suitable tool. Care must be taken so as not to
damage the pump housing or to incur further
damage to the contacts.
Inspection Recommendation
Deltec recommends annual functional inspec-
tion on the CADD-Prizm
®
pump. The follow-
ing inspection and testing procedures should be
performed annually to verify function and
accuracy.
NOTE:
Persons performing the following tests and
procedures should be familiar with the
Deltec CADD-Prizm
®
pump. Please read the
Operators Manual supplied with the pump
before proceeding.
WARNING:
CADD
®
pumps are sealed units. A broken or
damaged seal will, therefore, be considered
conclusive evidence that the pump has been
misused and/or altered, which voids any and
all warranties. All service and repair of
CADD
®
pumps must be performed by Deltec
or its authorized agents.
Cleaning
CAUTION:
Do not immerse the pump in cleaning
fluid or water. Do not allow solution to
soak into the pump, accumulate on the
keypad, or enter the battery compartment.
Do not allow solution to enter the data
in-out jack or the power jack. Make sure
the jack covers are closed before cleaning.
Do not expose the open air detector port
area of the pump or the connector
opening on the air detector to foreign
material, moisture or cleaning fluids. If
an air detector is not installed on the
pump, make sure the air detector port
cover is securely attached before cleaning.
Do not clean the pump with acetone,
other plastic solvents, or abrasive
cleaners.