1 Introduction
The Technical Manual is intended to provide
a basic, but limited, understanding of the
mechanical and electrical operation of the Deltec
CADD-Prizm
®
Computerized Ambulatory
Drug Delivery pump to persons familiar with
this device. The CADD-Prizm
®
Operator’s
Manual should be used in conjunction with
this publication for complete information.
This manual also outlines cleaning and func-
tional testing procedures that can be performed
on the CADD-Prizm
®
pump.
This technical manual is applicable to both the
CADD-Prizm
®
PCS and CADD-Prizm
®
VIP
pumps. For information on the CADD-Prizm
®
PCS pump refer to the information regarding
the CADD-Prizm
®
VIP PCA application in this
manual. Testing procedures in the back of this
manual are intended to be performed with the
pump in the PCA delivery mode.
IMPORTANT NOTICE:
CADD-Prizm
®
pump operations and safety
features are based on a microcomputer design.
Inadequate servicing or tampering with the
safety features of the pump may seriously
affect performance and safety.
For that reason, ALL SERVICING AND
REPAIR OF THE CADD-Prizm
®
PUMP
MUST BE PERFORMED BY DELTEC OR
ITS AUTHORIZED AGENTS.
The manufacturer’s warranty agreement shall
become null and void if the pump is not used
in accordance with the Operator’s Manual and
Instructions for Use for the pump accessories;
or, the pump is serviced by persons other than
Deltec or those authorized by Deltec.
Limited Warranty
The limited warranty associated with the
CADD-Prizm
®
pump can be found in the
product literature supplied with the product
when originally purchased, which is incorpo-
rated herein by reference. DELTEC SPECIFI-
CALLY DISCLAIMS ANY OTHER WAR-
RANTY, WHETHER EXPRESS, IMPLIED
OR STATUTORY, INCLUDING, WITHOUT
LIMITATION, ANY IMPLIED WARRANTY
OF MERCHANTABILITY OR FITNESS FOR
USE. Deltec further disclaims responsibility for
the suitability of the system for a particular
medical treatment or for any medical compli-
cations resulting from the use of the system.
The manufacturer shall not be responsible for
any incidental damages or consequential
damages to property, loss of profits, or loss of
use caused by any defect or malfunction of the
system.
If you wish to receive additional information
about the extent of the warranty on these prod-
ucts, please contact your Deltec representative or
call Customer Service at (800) 426-2448.
All recommendations, information and litera-
ture supplied by Deltec with respect to the
CADD
®
product line are believed to be accu-
rate and reliable, but do not constitute warran-
ties. No agent, representative, or employee of
Deltec has authority to bind Deltec to any
representation or warranty, expressed or
implied.
Exposure to Radiation or Magnetic
Resonance Imaging (MRI)
CAUTIONS:
(1) The pump SHOULD NOT BE
DIRECTLY IRRADIATED by therapeutic
levels of ionizing radiation because of the
risk of permanent damage to the pump’s
electronic circuitry. The best procedure to
follow is to remove the pump from the
patient during therapeutic radiation sessions
or diagnostic levels of radiographic and
fluoroscopic radiation. If the pump must
remain in the vicinity during a diagnostic or
therapy session, it should be shielded, and
its ability to function properly should be
confirmed following treatment.
(2) Magnetic fields produced by magnetic
resonance imaging (MRI) equipment may
adversely affect the operation of the pump.
Remove the pump from the patient during
MRI procedures and keep it at a safe
distance from magnetic energy.