Pump Cleaning and Functional Testing Checklist
The following checklist is provided as a guide only to assist in establishing documentation of cleaning and
functional testing for the CADD-Prizm
®
pump. If service is provided, fill out this sheet and return it with the
device.
Serial # _________________ Reference Number _________________________ Date _________________________
(Refer to the Technical Manual procedures.)
I. Cleaning Completed Yes No
II. Visual Inspection
LCD Lock Data In/Out Jack
Occlusion Sensor Seals Cassette Sensors (3) Air Detector or Port Cover
Valves and Expulsors Keyboard Pump Housing
Reservoir Hinge Area Indicator Light Battery Door
Latch Power Jack Battery Compartment
III. Mechanical Inspection
Keyboard Cassette Latch
Battery Door Cassette Lock
IV. Functional Inspection
Power-up RES VOL Alarm Dose Given / Attempted
Latch / Lock Stop / Start MG Given
Cassette Sensor Activation Timing Air Detector
LCD Dose Key Printer Tests
Motor / Gear Train Remote Dose Cord
V. Occlusion Tests (Only need to perform 1 or 2; not both)
Downstream Option 1: Activations Before Alarm ___________
Downstream Option 2: High Pressure Alarm At ____________ psi
Upstream Occlusion Sensor Test
Pass ________ Fail ___________
VI. Accuracy Testing (Only need to perform 1 or 2; not both)
Volumetric Accuaracy Test
Intended Actual Inaccuracy Accuracy Accuracy
Delivery Delivery Volume Error Error
Volume Volume Percentage
ml ml ml %
Gravimetric Accuracy Test
Pre-Delivery Post-Delivery Amount Intended Inaccuracy Accuracy Accuracy
Weights Weight Delivered Delivery Volume Error Error
Volume
g g ml ml ml %