NOTE:
The remaining testing procedures should be
performed using a 50- or 100-ml Medication
Cassette reservoir containing fluid and a
primed extension set with anti-siphon valve
or a primed CADD
®
administration set with
anti-siphon valve.
Starting/Stopping the Pump
Check the STOP/START key by pressing it.
Start the Pump? should be displayed. Press
the Y key. The display should show Starting
Pump followed by a review of the pro-
grammed parameters. The main screen
should appear with RUNNING in the
display, and the green LED indicator light
should blink every 3 seconds.
To stop the device, press the STOP/START
key. When the message Stop the Pump?
appears, press Y. STOPPED appears in the
display and the amber LED indicator light
blinks.
Activation Timing Check
Check the activation timing by programming
the pump with the following values:
Reservoir Volume: 1.0 ml
Units: Milligrams
Concentration: 1.0 mg/ml
Continuous Rate: 30.0 mg/hr
Demand Dose: 0.0 mg
Milligrams Given: 0.0 mg (press the
ENTER key to clear)
Press the STOP/START key. Press the Y key.
Starting Pump should appear on the
display. The pump should sequentially dis-
play all of the programmed values. Start a
timer at the first motor activation.
Count the activations. One activation should
occur every twelve seconds. Approximately
one minute fifty seconds (1:50) and ten
activations later, the RES VOL alarm should
occur. The display should show Reservoir
Volume is Zero with a Milligrams Given of
1.0 mg.
DOSE Key Check
Check the DOSE key operation by program-
ming the pump with the following values:
Reservoir Volume: 10.0 ml
Units: Milligrams
Concentration: 1.0 mg/ml
Continuous Rate: 0.0 mg/hr
Demand Dose: 1.0 mg
Demand Dose Lockout: 0 hrs 5 min
Max Doses Per Hour: 12
Dose Counters: 0/0 (Press the
ENTER key to clear)
Milligrams Given: 0.0 mg (Press the
ENTER key to clear)
Press the STOP/START key. Press the Y key.
The pump should sequentially display all of
the programmed values.
After RUNNING appears on the display,
press the DOSE key and note the time. The
pump should beep twice and begin to deliver.
Count the number of pump activations. The
pump should make ten double activations.
After ten double activations, the display
should show a reservoir volume of 9.0 ml.
Press the DOSE key two more times within
the next 5 minutes. The pump should not
deliver and the message Dose Not Deliv-
ered Dose Locked Out should be displayed.
Remote Dose Cord Check (if applicable)
Wait 5 minutes after the dose given above;
then, instead of pressing the DOSE key, press
the button on the Remote Dose cord. The
pump should make ten double activations.
After ten double activations, the display
should show a reservoir volume of 8.0 ml.
Press the DOSE key two more times within
the next 5 minutes. The pump should not
deliver and the message Dose Not Deliv-
ered Dose Locked Out should be displayed.
Doses Given and Doses Attempted Check
Stop the pump by pressing the STOP/START
key, then the Y key. Use the NEXT key to
advance to the Dose Counters screen. The
display should show 2/6, if the DOSE key
and Remote Dose cord were tested. If only
the DOSE key was tested, the display will
show 1/3. (If the above steps have not been
followed exactly, different values may appear.)