Deltech Fitness 6 1 0 0 Marine Sanitation System User Manual


 
6
Given
0 to 99999.99 in 0.01 unit increments
Clinician Bolus
0.1 ml to 20.00 ml (or mg or mcg equiva-
lent) (See Tables 2, 3 and 4 for scroll
ranges) Delivery rate (Continuous Rate +
Clinician Bolus): 125 ml/hr nominal
Continuous Delivery Mode Specifications
Reservoir Volume
1 to 9999 or Not In Use; programmable in
1 ml increments, displayed in 0.1 ml
increments
Default: 1 ml
Continuous Rate
0.1 to 350 ml/hr in the following increments:
0.1 for values between 0.1 and 100
1 for values between 100 and 350
Default: 0 ml/hr
Use 9-volt battery for rates up to 250 ml/
hr; use power pack or AC adapter for rates
up to 350 ml/hr
Given
0 to 99999.9 in 0.1 unit increments
TPN Delivery Mode Specifications
Reservoir Volume
10.0 to 9990 or Not In Use; program-
mable in 1 ml increments, displayed in 0.1
ml increments
Default: 10 ml
Infusion Volume
10 to 9990 ml in 10 ml increments:
Default: 10 ml
Infusion Period
0 hrs 10 min to 99 hrs 50 min in 10
minute increments:
Default: 1 hr 0 min
Taper-Up Period
0 hrs 0 min to 99 hrs 40 min in 10 minute
increments:
Default: 0 hrs 0 min
Taper-Down Period
0 hrs 0 min to 99 hrs 40 min in 10 minute
increments:
Default: 0 hrs 0 min
Plateau Rate
Calculated by pump; 10 to 350 ml/hr
Use 9-volt battery for rates up to 250 ml/hr;
use power pack or AC adapter for rates up to
350 ml/hr
KVO Rate
Calculated by pump: 1/10 of Plateau Rate up
to 5 ml/hr
Given
0 to 99999 in 0.1 unit increments
Intermittent Delivery Mode Specifications
Reservoir Volume
1 to 9999 or Not In Use; programmable in
1 ml increments, displayed in 0.1 ml increments
Default: 1 ml
Dose Volume
0.1 to 1000 ml in the following increments:
0.1 for values between 0.0 and 100
1 for values between 100 and 1000
Default: 0.0 ml
Dose Duration
1 min to 24 hrs in the following increments:
1 minute for values between 1 min and 10 min
5 minutes for values above 10 min
Default: 30 min
Duration is limited by Dose Volume so that
rate does not exceed 350 ml/hr. Use 9-volt
battery for rates up to 250 ml/hr; use power
pack or AC adapter for rates up to 350 ml/hr
Dose Cycle
10 min to 96 hrs in 5 minute increments:
Default: 4 hrs
KVO Rate
0 to 10 ml/hr in 0.1 ml/hr increments
Default: 0 ml/hr
Next Dose Start Time
10 min to 96 hrs X min (where X equals a 10
minute increment in the 96th hour) or Imme-
diate; programmable in 10 minute increments
Default: Immediate
Given
0 to 99999.9 in 0.1 unit increments
Time Remaining
Dose and Cycle display in 1 minute increments