Lock the device by inserting a key into the
lock and turning counterclockwise until the
mark lines up with the solid dot. The display
should show Cassette Locked.
Unlock the device by inserting a key into the
lock and turning clockwise until the mark
lines up with the open dot. The display
should show Cassette Unlocked.
Unlatch the reservoir by inserting a coin into
the latch slot and turning clockwise until the
mark lines up with the open dot. The display
should show Cassette Unlatched / Close
Clamp to Prevent Free Flow.
Cassette Sensor Check
Attach a 50- or 100-ml Medication Cassette
reservoir to the pump. Latch the cassette to
the pump. The display should show Reser-
voir latched. NOTE: The message displayed
depends on the type of reservoir or adminis-
tration set attached. See Table 13.
Lock the device by inserting a key into the
lock and turning counterclockwise until the
mark lines up with the solid dot. The display
should show Cassette Locked.
Unlock the cassette. The display should show
Cassette Unlocked. Unlatch the cassette.
The display should show Cassette Unlatched /
Close Clamp To Prevent Free Flow.
Functional Testing
The Power-up Check, Latch/Lock Check
and Cassette Sensor Check using a
Medication Cassette reservoir or CADD
®
administration set can be performed with the
pump in any of the four delivery modes.
The Cassette Sensor Check using the
CADD-Prizm
®
high volume administration
set can be performed in any delivery mode
except the PCA delivery mode.
Power-up Check
Insert a battery in the pump and observe the
LCD during power up.
The number values may vary depending on the current
mode and software revision.
If lec XXXXX (last error code with 5 digit
number) appears on the display prior to the
pump reviewing the current program settings,
the pump has experienced an electrical or
mechanical fault and should be returned for
service. If no error message is immediately
shown, the pump has powered up normally.
The pump should sequentially display all of
the programmed values. The words Self Test
Complete should appear, then the text
Power Up Successful with six audible
beeps. Continue with the Latch/Lock check.
Latch/Lock Check
Attach a 50- or 100-ml Medication Cassette
reservoir or a CADD
®
administration set to
the pump. The mark on the latch should be
aligned with the solid dot. The display should
show that the reservoir or administration set
is latched. See Table 13.
9 Testing Procedures
Reservoir Latched The Medication
Cassette reservoir has
been latched onto the
pump.
Admin set Latched The CADD
®
adminis-
tration set has been
latched onto the pump.
High Volume The CADD-Prizm
®
Admin Set Latched high volume adminis-
tration set has been
latched onto the pump.
Table 13.
(excluding PCA
delivery mode)
C A D D - P R I Z M 61XX
lec XXXXX
sn XXXXXXXX 0254-01X
625 XXXXX XXXX-X
Terminal Software revision level
Error code Pump model number
Serial number