Deltech Fitness 6 1 0 0 Marine Sanitation System User Manual


 
5
Specifications (Nominal)
General Pump Specifications
Resolution
Medicaton Cassette reservoir or CADD
®
administration set, 0.050 ml/pump stroke
nominal
CADD-Prizm
®
high volume administration
set, 0.100 ml/pump stroke nominal
Size
4.4 cm x 10.4 cm x 14.1 cm (1.7 in. x 4.1
in. x 5.6 in.) excluding cassette or other
accessories
Weight
568 g (20 oz.) including 9-volt battery and
empty 100-ml Medication Cassette reservoir,
excluding other accessories
Pump Alarms
Low battery power; depleted battery power;
external power source low, faulty, depleted;
pump stopped; pump fault; low reservoir
volume; high delivery pressure; air in line;
air detector faulty or detached (only with the
use of the optional air detector); air detector
port cover detached; delivery too slow; key
stuck; cassette detached or unlocked; print
failure, upstream occlusion (model 6101 only)
Bolus Volume at Occlusion Alarm Pressure
0.050 ml resolution sets/reservoirs:
< 0.25 ml
0.100 ml resolution sets: < 2.0 ml
Power Sources
9-volt alkaline or lithium battery such as
DURACELL
®
Alkaline MN 1604 or
ULTRALIFE
®
Lithium U9VL; Power Pack
reorder number 21-3801; AC adapter.
An internal battery powers the clock. When
it is depleted, it cannot reliably maintain the
clock time. This battery must be replaced by
the manufacturer. The internal battery has
an expected life of 5 years.
System* Operating Temperature
+2°C to 40°C (36°F to 104°F)
System* Storage Temperature
-20°C to 60°C (-4°F to 140°F)
Power Pack Charging Temperature
+10°C to 35°C (50°F to 95°F)
System* Delivery Accuracy
± 6% (nominal)
High Pressure Alarm
18 ± 9 psi
Air Detector Alarm
Single bubble greater than 0.100 ml
PCA Delivery Mode Specifications
Reservoir Volume
1 to 9999 or Not In Use; programmable in
1 ml increments, displayed in 0.1 ml increments
Default: 1 ml
Units
Milliliters (ml), milligrams (mg), micrograms (mcg)
Default: milligrams
Concentration
Mg/ml: 0.1, 0.2, 0.3, 0.4, 0.5, 1, 2, 3, 4, 5, 10,
15, ... 95, 100
Mcg/ml: 1, 2, 3, 4, 5, 10, 15, ...95, 100, 200,
300, 400, 500
Default: 1 mg/ml
Continuous Rate
0 to 30 ml/hr (or the mg or mcg equivalent)
(See Table 1 for scroll ranges)
Default: 0 mg/hr
Demand Dose
0 to 9.9 ml (or the mg or mcg equivalent)
(See Tables 2, 3 and 4 for scroll ranges)
Default: 0 mg/hr
Delivery rate (Continuous Rate + Demand
Dose): 125 ml/hr nominal
Demand Dose Lockout
5 minutes to 24 hours in the following increments:
1 minute for values between 5 and 20 minutes
5 minutes between 20 minutes and 24 hours
Default: 5 minutes
Max Doses Per Hour
1 to 12 doses in 1 dose increments (will also be
limited by the Demand Dose Lockout value)
Default: 1
Demand Doses Given
0 to 999
Demand Dose Attempts
0 to 999
*System is defined as a CADD-Prizm
®
pump with an attached Medication Cassette reservoir and CADD
®
extension set with integral
anti-siphon valve, or an attached CADD
®
administration set with integral or add-on anti-siphon valve.