Micronics 216 Marine Sanitation System User Manual


 
13
Sensor Parameters
This facility is password protected. It stores sensor information used by Micronics and is not available for the user.
WARNING! Sensor
should only be edited
following instruction
from the factory
Enter password
Factory Settings
The facility is used by Micronics in the process of instrument calibration. Press ENTER takes you back to
SETUP
INSTRUMENT MENU
.
Exit
Means EXIT and will take you back to the Main Menu.
Main Menu - Read Flow
When choosing the
Read flow
option from the
MAIN MENU
the instrument reverts directly back to the data that was
last entered. The instrument will have to be reprogrammed if it is to be used on a new application.
KEYPAD OPTIONS
The output options can only be adjusted/operated in flow mode.
Exit
Delete Key
If anything is entered in error, press the DELETE key and re-enter the information required.
Options Key
This can only be used in flow mode. Scroll down the options then press
ENTER
to select.
OPTIONS
Cutoff (m/s) 0.05
Set zero
Total RUN
Reset + total
Reset – total
Damping (sec) 5
Cal Factor 1000
Corr Factor 1000
Diagnostics
Exit
Cut Off (m/s)
The instrument has an automatic CUTOFF that is calculated to 0.05 m/s. The maximum flow is calculated when the
instrument is programmed and is displayed when sensor set and mode of operation are displayed (See page 9 - Read
Flow - Attach sensors). Micronics cannot guarantee measuring flows below this range because of instabilities in
measuring system, but it is possible for the user to cancel any cut-off altogether.
Reducing the cutoff to Zero allows the user to see or record any flow that they may not want. For example it may be
that the user may not want to measure flows below 50 LPM in a 50mm pipe that is equivalent to 0.42 m/sec, in which
case 0.42 m/sec would be entered into the instrument and nothing would be recorded below that level. The maximum
cut off
is 1 m/sec.