Micronics 216 Marine Sanitation System User Manual


 
9
The display will now show the sensor seperation distance. Adjust the moveable sensor to the required distance. Press
ENTER
to read flow.
READ FLOW
now appears on the display.
Batt CHRG Sig 48%
(ERROR MESSAGES APPEAR HERE)
m/s
When reading volumetric flow the instrument will display a positive and negative total. Selecting OPTIONS from the
keypad can reset these totals (See page 14).
The instrument will continually display the battery and signal levels. Signal levels should be above 40%.
If there is an error with the site data entered or the application, the instrument will display an Error or warning message
(See page 15-17), which will appear above the flow reading. If there is more than one message it will scroll between
them all.
To stop reading flow press ENTER
ONCE
. The display will read the following.
This will stop all
outputs
Press ENTER to EXIT
SCROLL to return
to READ FLOW
Pressing ENTER a second time will stop outputs and return the instrument to
MAIN
MENU or
Press the scroll key to
return the instrument to
READ FLOW
.
Main Menu - View/Edit Site Data
The
VIEW/EDIT SITE DATA
mode can be accessed from the main menu. It allows the user to enter application
details for up to 20 different sites. This facility is useful if a number of sites are being monitored on a regular basis.
Application data can be programmed into each site before getting to site.
When scrolling up/down the menu press
ENTER
to select at each prompt.
VIEW/EDIT SITE DATA
List sites
Site number 0
Name QUICK START
Units MM
Set sensor
Separation to
XXX
ENTER to continue