3M SE-400 SONAR User Manual


 
SoundPatrol feature (non-IS models only)
Printing and reports
Sound Examiner SE-400 Series
27
Printing and reports
The SoundPatrol features (non-IS models only) an easy to access print icon , located on the
measurements and calibration screens, which is used for report printing when connected to the handheld
printer (sold as an option). The following three reports are explained below.
SoundPatrol Reports and Samples
Description
1-Second Continuous Data Report:
TResponse = SLOW
Time, SPL, Leq, Weight = C
2, 68.0, 67.4
3, 69.0, 68.0,
4, 70.1, 68.4,
5, 71.3, 69.2
6, 71.2, 69.8,
7, 68.1, 69.8,
3M SE-400 R.10D
S/N: SEK06BETA06
The 1-second continuous data report provides a printout of 1- second
measurements of the sound pressure level values and the average
(Leq) values at the appropriate time response (fast/slow) and frequency
weighting (A, C, or Z).
To print the 1-second continuous data report, power on the SE-400
Series and press the key while the printer is connected. Once
in run mode, press the softkey from the measurement screen
and the connected hand-held printer will print the report. At any
time, press icon to end the printing. The handheld printer will
stop the 1-second report.
Summary Data Report:
TResponse = SLOW
SPLmax = 78.8 dBA
Leq = 71.5 dBA
L
AE
= 84.9 dBA
SPLmin = 64.4 dBA
Runtime = 00:00:17
Date: 15-Mar-2012
Time: 9:21:54
Location:__________________
By: ______________________
The summary data report displays the time response and the summary
data parameters including the maximum sound pressure level (SPL),
the average SPL (Leq), the minimum SPL, the runtime, date, time,
location, and a by field (for an optional signature). These parameters
may be changed via the print setup menu. (See Print setup section
above.)
Note: this report will print after the 1-second continuous data report
has finished printing (or is stopped).
To print the summary data report, press key on the SE-400
while the printer is connected. Press the softkey. Press
on the SE-400 and the summary data report will appear after the 1-
second continuous report.
Calibration Report:
3M SE-400 R.10D
S/N: SEK06BETA06
TResponse = SLOW
CALIBRATION DATE:
14:21:05 15-MAR-12
CALIBRATION INFO:
Old SPL = 113.8 dBA
Ref Level = 114.0 dBA
New SPL = 114.0 dBA
The calibration report displays the time response setting, the calibration
date and time, the previous SPL (noted as old), the reference level
value, and the new calibrated level.
To print the calibration report, press the softkey from the
calibration screen while connected to the printer. To access from a
measurement screen, press the icon and then press the
CAL softkey.