3M SE-400 SONAR User Manual


 
Data Addendum
Sound Examiner SE-400 Series
60
Pressure to free field corrections with BK4936 microphone and RICR, windscreen, and remote preamp
(sections: 5.2.8, 9.2.7d)
Note: add to pressure response to get 0incidence free-field response. B&K 4226 calibrator may be used to determine the
pressure field response.
Frequency
in Hz
BK4936 microphone
Pressure Field to
Free Field
Corrections
in dB
32
0.1
40
0.0
50
-0.1
63
-0.1
80
-0.1
100
-0.1
125
-0.1
160
-0.1
200
-0.2
250
-0.1
315
0.0
400
0.1
500
0.2
630
0.1
800
0.0
1000
0.0
1250
0.2
1600
0.6
2000
0.9
2500
1.2
3150
1.8
4000
2.8
5000
3.5
6300
5.3
8000
7.1
10000
9.2
12500
9.3
16000
9.1