3M SE-400 SONAR User Manual


 
Data Addendum
Sound Examiner SE-400 Series
40
Data Addendum: SE-400 Series measuring to IEC 61672-1
About this data addendum
This data addendum presents submission information for testing the 3M
TM
Sound Examiner SE-400 Series sound level meter
to the IEC61672-1 standard. The SE-400 Series is a data logging sound level meter.
Note: The nominal microphone sensitivity is -28 dB.
Linear operating range (sections: 5.5.10, 5.5.11, 9.2.7e, 9.3e, 9.3f)
The starting level for measuring level linearity is 114 dB at all frequencies. Calibration point is 114 dB at 400 mV @ 1 kHz.
Linear operating range with SPL A-Weighting and LEQ A-Weighting (sections: 5.5.10, 5.5.11, 9.2.7e, 9.3e, 9.3f)
Note: Reference level 114 dB, Cal Point: 400mVrms = 114 dB @ 1 kHz
Frequency
SPL A-Weighting
LEQ A-Weighting
Linear Operating Range (dB)
Linear Operating Range (dB)
31.5
75
140
75
140
1000
34
140
34
140
4000
33
140
33
140
8000
37
140
37
140
12500
42
140
42
140
Linear operating range with SPL C-Weighting and LEQ C-Weighting (sections: 5.5.10, 5.5.11, 9.2.7e, 9.3e, 9.3f)
Note: Reference level 114 dB, Cal Point: 400mVrms = 114 dB @ 1 kHz
Frequency
SPLC-Weighting
LEQ C-Weighting
Linear Operating Range (dB)
Linear Operating Range (dB)
31.5
43
140
43
140
1000
38
140
38
140
4000
39
140
39
140
8000
42
140
42
140
12500
47
140
47
140
Type/Class 1 and Type/Class 2
Equivalent input
voltage
0.400 VRMS
Capacitance
12 pF
Resistance
>1G