Teledyne 450 Marine Instruments User Manual


 
6. Turn on the M450 by switching the switch on the lower right corner of the
front panel. The front panel display should light with a sequence of messages,
-API - M450 - software version number, then a normal display as shown in
Figure 2.2.
7. Allow about 20 minutes for the UV lamp temperature to come up to it’s
setpoint then press the TST> button on the front keyboard to scroll through
the TEST values. Compare these values to those noted during the final
factory checkout listed in Table 1.2. The values observed should closely
match the Table 1.2 values.
CAUTION
Verify that the instrument is set up for proper line voltage
and frequency.
NOTE
Words in all caps are messages on the monitor front
panel.
FIGURE 1.1 - REAR PANEL PNEUMATIC CONNECTIONS
P/N 02395D1 Teledyne API Model 450 O
3
Monitor Instruction Manual - Page 13