3.0 SETUP MODE
This section describes the setup variables which are used to configure the Monitor. All
the setup variables are stored in the Monitor's EEPROM and are retained during power
off and even when new software revisions are installed.
NOTE
If a variable is modified, but ENTR is not pressed, the variable will
not be changed and the monitor will beep when exit is pressed.
The setup menus are accessed by pressing the SETUP button from the instrument’s main
menu. The top level setup menu is shown in Figure 3.1 below.
SEQ
STRM UNIT COMMCLK VARS DIAG EXIT
SETUP A.8
FIGURE 3.1 - SETUP MENU
3.1 Configuring Auto Sequence (SEQ)
The automatic sequence determines the monitoring order for the streams in a multi-
stream instrument. The default setting for a six channel instrument is 1-2-3-4-5-6. With
this setting the instrument will cycle through all six channels in numerical order and loop
back to the beginning when the cycle is complete. This sequence can be modified,
however, to cycle through the channels in any order desired. The sequence can also be
modified to monitor certain streams more often than others, 1-2-1-3-1-4-1-5-1-6 for
example.
To modify the auto sequence, press SEQ from the setup menu. The sequence editing
screen is shown in Figure 3.2 below.
P/N 02395D1 Teledyne API Model 450 O
3
Monitor Instruction Manual - Page 39