Teledyne 465L Marine Instruments User Manual


 
05509F DCN6409 15
2. SPECIFICATIONS AND AGENCY
APPROVALS
2.1. Specifications
Ranges
User selectable to any full scale range from:
1 PPM to 500 PPM
1000 PPB to 10,000 PPB
2000 µg/m
3
to 10,000 µg/m
3
100 PPHM to 10,000 PPHM (optional
configuration)
Measurement Units
PPM, PPB, µg/m
3
, PPHM (optional
configuration)
Zero Noise < .0015 PPM (rms)
Span Noise < .5% of reading (rms) (above 0.1 PPM)
Lower Detectable Limit < .003 PPM (rms)
Accuracy +/- 1% of full scale range
Response Time (95%)
<30 sec
(<60 sec with the Sample Conditioner option)
Sample Flow Rate 0.8 L/min
Temperature Range 5-45
o
C
Humidity Range 10-90% RH, Non-Condensing
Dimensions (H x W x D)
5.22” x 19.0" x 15.3" (133 mm x 483 mm x 388
mm) – Rack Mount Version
16.01” x 15.6” x 6.90” (407mm x 396mm x
175mm) – with NEMA 4X enclosure
Weight
12.6 lb. (5.73 kg) - Rack Mount Version
15.2 lb. (6.91 kg) – with NEMA 4X enclosure
Power 100-240 VAC, 47-63 Hz, 74W max
Environmental Conditions
Installation Category (Overvoltage Category) II
Pollution Degree 2
Maximum Operating Altitude 2000 meters