Teledyne 465L Marine Instruments User Manual


 
Teledyne API Model 465L O
3
Monitor Operation Manual
05509F DCN6409 53
7.2.7. SETUP-STREAM Menu
The STREAM menu allows the operator to view and edit properties of each measurement stream
available. For each stream the following properties are available:
Table 7-8: STREAM Properties
Property Meaning Default
Value
Range
DURATION Stream measurement duration for
AUTO stream-selector operation.
See Section 7.3 for additional
details.
60.0 Seconds 60.0 – 360.0
Seconds
HI ALARM HI alarm trigger value 0.100 PPM .010 – 9.999
PPM
*
HIHI ALARM HI-HI alarm trigger value 0.300 PPM .010 – 9.999
PPM
*
ON/OFF HI
ALARM
HI alarm enable, enables HI alarm
operation
ON ON/OFF
ON/OFF HI HI
ALARM
HI-HI alarm enable, enables HI-HI
alarm operation
ON ON/OFF
RANG Concentration range for specified
stream. Used to scale the analog
output.
1.0 PPM 1.0 – 500.0
PPM
ENABLE Allows the choice of viewing or
skipping each stream in a multi-
channel instrument when cycling
through the streams in Auto or
Manual modes. OFF = skip. At
least one stream must be set to ON
(enabled).
ON ON-OFF
*
For low concentration instrument ranges. For certain range configurations, the
alarms may be set to higher values