7.2.4. VIEW Menu
The VIEW menu allows the operator to view various measurement parameters in real-time. This
can be useful for diagnosing various instrument or system problems.
Note that these values are updated in real-time as they are repeatedly scrolled on the display.
Table 7-5: VIEW Menu Parameters
Parameter Meaning Units Normal Range*
MEAS UV detector reading, measure cycle mV 250 – 1230
REF UV detector reading, reference
cycle
mV 250 – 1230
PRESSURE Sample pressure psia 9.0 – 14.9
LAMP
TEMP
UV lamp temperature ºC 50.0 – 52.0
SAMPLE
TEMP
Sample temperature ºC 5.0 –50.0
FLOW Sample Gas Flow Rate cc/min 640 – 960
*After initial 20 minute warm-up period.
7.2.5. CONFIG Menu
The CONFIG menu displays specific information related to the instrument’s configuration. No
changes are made within this menu.
Table 7-6: CONFIG Menu Parameters
CONFIG Meaning
S/N Serial number of M465L. (Should
match the serial number tag on the rear
panel of the instrument).
COMM ADDRESS Displays the MODBUS slave ID for
the instrument.
STREAM COUNT Displays the number of streams
configured. (Always 1 for instruments
without the Multi-stream Option).
REV Revision of instrument firmware
BREV Revision of sensor firmware
BOOT Revision of bootloader firmware