8.6.3. Real-Time Concentration and Status Registers (Read
Only)
Decimal
Address Description Register Type Start Bit # Bits
1 Stream 1 Concentration (PPB) Float Inverse 0 32
3 Stream 2 Concentration (PPB) Float Inverse 0 32
5 Stream 3 Concentration (PPB) Float Inverse 0 32
7 Stream 4 Concentration (PPB) Float Inverse 0 32
9 Stream 5 Concentration (PPB) Float Inverse 0 32
11 Stream 6 Concentration (PPB) Float Inverse 0 32
13 Stream 1 HI Alarm Status, 1= ON Byte 0 8
13 Stream 2 HI Alarm Status, 1= ON Byte 8 8
14 Stream 3 HI Alarm Status, 1= ON Byte 0 8
14 Stream 4 HI Alarm Status, 1= ON Byte 8 8
15 Stream 5 HI Alarm Status, 1= ON Byte 0 8
15 Stream 6 HI Alarm Status, 1= ON Byte 8 8
16 Stream 1 HI-HI Alarm Status, 1= ON Byte 0 8
16 Stream 2 HI-HI Alarm Status, 1= ON Byte 8 8
17 Stream 3 HI-HI Alarm Status, 1= ON Byte 0 8
17 Stream 4 HI-HI Alarm Status, 1= ON Byte 8 8
18 Stream 5 HI-HI Alarm Status, 1= ON Byte 0 8
18 Stream 6 HI-HI Alarm Status, 1= ON Byte 8 8
19 Stream 1 Active Status, 1= Sampling Byte 0 8
19 Stream 2 Active Status, 1= Sampling Byte 8 8
20 Stream 3 Active Status, 1= Sampling Byte 0 8
20 Stream 4 Active Status, 1= Sampling Byte 8 8
21 Stream 5 Active Status, 1= Sampling Byte 0 8
21 Stream 6 Active Status, 1= Sampling Byte 8 8
24 Gas Flow (cc/min) Float Inverse 0 32
39 Sensor Reference Value (mV) Float Inverse 0 32
41 Sensor Measure Value (mV) Float Inverse 0 32
43 Sensor Pressure Value (psia) Float Inverse 0 32
45 Sensor Gas Temperature (K) Float Inverse 0 32
47 Sensor UV Lamp Temp (°C) Float Inverse 0 32