The last Check Strip
reading was outside the
acceptable range of
results. The Check Strip
test must be repeated. (See Chapter 3,
Checking the System.) If the message
continues, clean the Meter. (See Chapter
8, Cleaning and Maintaining the Meter.)
For further assistance, call the LifeScan
Technical Service Department for more
information.
Usually indicates a test
done with Control
Solution. However, if this
message appears after a
blood test, it means that
your blood sample was
too small, smeared, or a second drop was
applied after the test began. Your test
result will not be accurate. Repeat the
test using a larger drop of blood. (See
Check the Amount of Blood on the Test
Strip in Chapter 6, Testing Your Blood.)
All segments on the
display will appear when
the On/Off Button is pressed and held. If
any part of the display does not appear,
call the LifeScan Technical Service
Department: