Emerson Rosemount 3051CF Marine Instruments User Manual


 
Quick Start Guide
11
June 2014
Step 4: Set the switches
Set Simulate and Security switch configuration before installation as shown in
Figure 8.
The simulate switch enables or disables simulated alerts and simulated AI
Block status and values. The default simulate switch position is enabled.
The Security switch allows (unlocked symbol) or prevents (locked symbol)
any configuration of the transmitter.
- Default security is off (unlocked symbol).
- The security switch can be enabled or disabled in software.
Use the following procedure to change the switch configuration:
1. If the transmitter is installed, secure the loop, and remove power.
2. Remove the housing cover opposite the field terminal side. Do not remove the
instrument cover in explosive atmospheres when the circuit is live.
3. Slide the security and simulate switches into the preferred position.
4. Replace the housing cover.
Note
It is recommended the cover be tightened until there is no gap between the cover and the housing.
Figure 8. Simulate and Security Switches
A. Simulate Disabled Position
B. Simulate Switch
C. Simulate Enabled Position (default)
D. Security Locked Position
E. Security Switch
F. Security Unlocked Position (default)
C
A
B
D
E
F
00825-0100-4774_RevFA.fm Page 11 Friday, June 6, 2014 2:21 PM