Quick Start Guide
DP Level example
Display pressure on the LCD display
Select the 'pressure' check box on the display configuration screen.
Step 7: Zero trim the transmitter
Note
Transmitters are shipped fully calibrated per request or by the factory default of full scale (span =
upper range limit).
A zero trim is a single-point adjustment used for compensating mounting
position and line pressure effects. When performing a zero trim, ensure that the
equalizing valve is open and all wet legs are filled to the correct level.
The transmitter will only allow 3-5% URL Zero error to be trimmed. For greater
zero errors, compensate for the offset by using the XD_Scaling, Out_Scaling and
Indirect L_Type which are part of the AI Block.
1. To use guided setup:
a. Navigate to Configure, Guided Setup, and select "Zero Trim".
b. The method will execute the zero trim.
2. To use manual setup:
a. Navigate to Overview, Calibration, Sensor Trim, and select "Zero Trim".
b. The method will execute the zero trim.
Parameters Enter data
Channel 1
L_Type Indirect
XD_Scale 0 - 300 inH
2
0 @ 68 °F
Note
Select only the units that are supported by the device.
Out_Scale 0-25 ft.