Emerson Rosemount 3051CF Marine Instruments User Manual


 
Quick Start Guide March 2014
China
E3 China Flameproof
Certificate: GYJ14.1041X; GYJ10.1313X [Flowmeters]
Standards: GB3836.1-2000, GB3836.2-2010, GB12476-2000
Markings: Ex d IIC T6/T5, T6(-50°C Ta +65°C), T5(-50°C Ta +80°C)
Special Conditions of Use (X):
1.
The relation between ambient temperature arrange and temperature class is as follows:
Ta Temperature Class
-50°C~+80°C T5
-50°C~+65°C T6
When used in a combustible dust environment, the maximum ambient temperature is
80°C.
2.
The earth connection facility in the enclosure should be connected reliably.
3.
Cable entry certified by notified body with type of protection Ex d IIC in accordance wi
th
GB3836.1
-2000 and GB3836.2-2000, should be applied when
installed in a hazardous
location. When used in combustib
le dust environment, cable entry in accordance wi
th
IP66 or highe
r level should be applied.
4.
Obey the warning “Keep tight when the circuit is alive.”
5.
End users are not permitted to change any internal
components.
6.
During installation, use and maintenance of this product, observe the followi
ng
standards:
GB3836.1
3-1997 “Electrical apparatus for explosive gas atmospheres Part 13: Repai
r
and overhaul fo
r apparatus used in explosive gas atmospheres”
GB3836.15-2000 “Electrical apparatus for explosive gas atmospheres Part 15: Electr
ical
installations
in hazardous area (other than mines)”
GB3836.16-2006 “Electrical apparatus for explosive gas atmospher
es Part 16:
Inspection and maintenance of ele
ctrical installation (other than mine
s)”
GB50257
-1996 “Code for construction and acceptance of electric dev
ice for explosion
atmospheres an
d fire hazard electrical equipment installation engineering”.
GB12476.2-2006 “Electrical apparatus for use in the presence of co
mbustible dust Part
1-2: Electrical apparatus prot
ected by enclosures and surface temperature limitation-
Selection, installation and maintenance”
GB15577-2007 “Safety regulations for dust explosion prevention and protection”
I3 China Intrinsic Safety
Certificate: GYJ13.1362X; GYJ101312X [Flowmeters]
Standards: GB3836.1-2010, GB3836.4-2010, GB3836.20-2010, GB12476.1-2000
Markings: Ex ia IIC Ga T4/T5
Special Conditions for Safe Use (X):
1.
Symbol “X” is used to denote specific conditions of
use:
a.
If the apparatus is fitted with an optional 90V trans
ient suppressor, it is not capable
of
withstanding the 500V insulation test for 1 minute. This must be ta
ken into
account w
hen in
stalling the apparatus.
b.
The enclosure may be made of aluminum alloy and given a protective polyuretha
ne
paint finish; however, care should
be taken to prot
ect it from impact or abrasion if
located in Zone 0
.
2.
The relation between T code and ambient temperature ran
ge is:
Model T Code Temperature Range
HART T5 -60°C Ta +40°C
HART T4 -60°C Ta +70°C
Fieldbus/PROFIBUS/FISCO T4 -60°C Ta +60°C
Flowmeter with 644 Temp Housing T4 -40°C Ta +60°C
3. Intrinsically Safe parameters:
HART Fieldbus/PROFIBUS FISCO
Voltage U
i
30 V 30V 17.5 V
Current I
i
200 mA 300 mA 380 mA