3.2 Loading QC Setup and Reference Entry
The 978 will store a separate reference for each of 30 Setups. The
reference will remain with the selected "Setup" until changed.
When the "AUTO SETUP" option in Sec. 7.1 is set to "OFF", references
can be either measured or manually entered. When "AUTO SETUP" is set to
"ON", references can only be measured. A reference is set after the Setup is
"loaded" into the display.
Loading a QC Setup
Make sure that either "PASS/FAIL", 555 SHADE TAG", or "555 & PASS/FAIL" function is
selected before attempting to load a Setup. Refer
to Section 1.5 for function select information if
required.
1) Repeatedly press [SETUP] to select
the setup # to load.
There is approx. 3 seconds allowed between
depressions of the "SETUP" key before the
display returns to normal operation.
Setup #’s can be decremented by pressing and
holding the [SETUP] key, then pressing the
[FUNCTION] key.
2) After desired setup is displayed,
press [FUNCTION] (load) key to load setup.
Reference Measurement
After the desired Setup is loaded, a
reference can be set.
1) Press the [REF] key to enter
reference function.
"SET REFERENCE" is display then
the reference menu.
- Continue with Step 2 to manually
enter reference values, or advance
to Step 3 to measure a reference.
3-10
SETUP # 3
FUNCTION SETUP REF
load
define
PRESS REPEATEDLY
SETUP # 3
LOADED!
E .00
FUNCTION SETUP REF
a* +.00
SET REFERENCE
SELECT
exitenter read
L* +.00
b* +.00