Pass/Fail and 555 Shade Tag Measurements . . . continued
To take a QC measurement:
Defining Setups (Sec. 3.1) must be performed before continuing with this procedure.
If QC measurement are to be Stored in the instrument, refer to Section 5 for Storage selection
procedure.
The following procedure will guide you through all the steps required to take
a setup measurement. Measurement tagging using the optional bar code reader
will be covered in Tagging and Storing Operation, Section 5.4.
1) Make sure "PASS/FAIL", "555
SHADE TAG", or "555 & PASS/FAIL"
is selected. Repeatedly press
[SETUP] key to select setup.
2) Press [FUNCTION] (load) key to load
the selected Setup.
The [FUNCTION] (load) key must be pressed
within 3-sec. or the display will revert back to
operation mode.
If a reference does not exist for the selected
setup, refer to Section 3.2 for measurement
procedure.
3) Press the [FUNCTION] key to select
"PASS/FAIL", "555 SHADE TAG", or
"555 & PASS/FAIL"
The display function can be changed after a
measurement is taken to view sample data in a
different function.
3-14
SETUP # 1
FUNCTION SETUP REF
CAL
STORE
(-)
(+)
( )
load
define
PRESS REPEATEDLY
SETUP # 2
FUNCTION SETUP REF
load
define
SETUP # 2
LOADED!
FUNCTION =
FUNCTION SETUP REF
PASS / FAIL
PASS/FAIL
555 SHADE TAG
555 & PASS/FAIL