Monitor Instruction Manual - Page 20
Figure 4-4: Pneumatic Connections, NEMA Configuration
4.4.1. Ozone Inlet Connection
Sample gas connections are made to the ¼” stainless steel compression fitting on the rear
panel. (See Figure 4-3 and Figure 4-4)
1/4" O.D. FEP (Teflon) or Stainless Steel tubing is recom
mended to connect the sample
source to the monitor. Any fittings used in the sample lines should be constructed of
Stainless Steel or Teflon.
The M465M has an internal sample pump and is designed to sample gas that is close to
ambient pressure. Ensure that the sample gas pressure is in the range of 11 – 16 psia
(pounds per square inch absolute.)