Figure 7-4: Span Cal Menu
7.2.4. VIEW Menu
The VIEW menu allows the operator to view various measurement parameters in real-
time. This can be useful for diagnosing various instrument or system problems.
Table 7-5: VIEW Menu Parameters
Parameter Meaning Units Normal Range*
MEAS UV detector reading, measure cycle mV 250 – 1230
REF
UV detector reading, reference
cycle
mV 250 – 1230
PRESSURE Sample pressure psia 9.0 – 18.0
LAMP
TEMP
UV lamp temperature ºC 50.0 – 52.0
SAMPLE
TEMP
Sample temperature ºC 5.0 –50.0
FLOW Sample Gas Flow Rate** ON/OFF ON
*After initial 20 minute warm-up period.
** Optional Hardware.
NOTE
These values are updated in real-time as they are repeatedly
scrolled on the display.