Teledyne 465M Marine Radio User Manual


 
P/N 06160D DCN6409 Teledyne API Model 465M O
3
Monitor Instruction Manual - Page 51
9. MAINTENANCE
9.1. Maintenance Schedule
Table 9-1 below outlines the suggested maintenance procedures and intervals for
ensuring the M465M continues to operate accurately and reliably.
Table 9-1: Maintenance Schedule
Maintenance Item Recommended Interval Section
Replace particulate filter
element
6 months
*1
9.3
Adjust UV lamp As Indicated by ‘Check Lamp’
LED or status output
9.4
Replace lamp As required; when adjustment can
no longer be performed.
9.5
Replace sample pump 2 years 9.6
Replace Auto-Zero Valve 2 years 9.7
Rebuild cells, replace o-
rings
3 years Return
to
factory
*1
When external sample line pre-filters are used. If pre-filters are not
used, internal sample filter should be replaced every month.
CAUTION
RISK OF ELECTRICAL SHOCK. THE OPERATIONS
OUTLINED IN THIS CHAPTER ARE TO BE PERFORMED BY
QUALIFIED MAINTENANCE PERSONNEL ONLY!