The firmware loader software should display a message “Boot loader found,
verifying…”. If this message is not displayed, power down the instrument and
receck the COM port configuration and cable (see Section 9.10.1) and retry.
3. When the upload is complete, a message, “Verify-flash completed successfully”
should be displayed. If it is an unsuccessful upload, a message “Verify-flash
error” is displayed. If a Verify-flash error occurs, ensure that the correct PC
COM port is selected for the instrument (see Section 9.10.2, step 4).
NOTE
Search for boot loader will time out in 20 seconds. If the
instrument is not powered up in 20 seconds, the screen will
display, “No boot loader detected.” Go back to step 1 to verify
the firmware.
NOTE
A verification process can be stopped at any point in time. This
will not affect the firmware that is already uploaded on the
instrument. The instrument will automatically go to the firmware
in 10 seconds.