Table 5-5: GPT and GPT Preset Step Properties
Property Description
O3 Conc. The numeric concentration of O
3
gas to generate.
O3 Units The O
3
concentration unit to use. The M700 will perform any necessary unit
conversions.
NO Conc. The numeric concentration of NO gas to generate.
NO Units The NO concentration units to use.
5.4.9. Manual Steps
A Manual step will display the following property dialog. The properties are described in Table 5-6.
Figure 5-11: Manual Step Properties Dialog
Table 5-6: Manual Step Properties
Property Description
Cal. Flow Flow rate of calibration gas (LPM - liters per minute).
Cal. Gas The gas type to generate. Although APIcom permits to specify several different
gases, the M700 may not be able to generate them all. If the M700 cannot generate
a specified gas, it will simply generate zero air.
Dil. Flow Flow rate of the diluent gas (LPM - liters per minute).
O3 Generator
Mode
For manual GPT. If Constant Drive or Reference Feedback are selected, the units
for O
3
are mV. If Bench Feedback is selected, the units for O
3
are PPB.
O3 Setpoint The mV or PPB setting for O
3
.
5.4.10. Duration Steps
Property dialog box & properties table.
Figure 5-12: Duration Step Properties Dialog