Emerson PN 51-FCL-1056 Marine Instruments User Manual


 
20
MODEL FCL-1056 SECTION 5.0
PROGRAMMING THE ANALYZER
TABLE 5-1. DEFAULT SETTINGS
ITEM CHOICES DEFAULT
Sensor assignment
1. Sensor 1
chlorine chlorine
2. Sensor 2
pH pH
Outputs
1. Assignments (if FCL-01)
a. output 1
chlorine, temp chlorine
b. output 2
chlorine, temp temp
2. Assignments (if FCL-02)
a. output 1
chlorine, pH, temp chlorine
b. output 2
chlorine, pH, temp pH
3. Range 0-20 or 4-20 mA 4 – 20 mA
4. 0 or 4 mA setting
a. chlorine and pH
-9999 to +9999 0
b. temperature
-999.9 to +999.9 0
5. 20 mA setting
a. chlorine
-9999 to +9999 10
b. pH
-9999 to +9999 14
c. temperature
-999.9 to +999.9 0
6. Fault current (fixed) 0.00 to 22.00 mA 22.00 mA
7. Dampening
0 to 999 sec 0 sec
8. Simulate
0.00 to 22.00 mA 12.00 mA
Alarms
1. Logic
high or low AL1 low, AL2,3,4 high
2. Assignments
a. AL1 and AL2
chlorine, pH, temp, fault,
interval timer
chlorine (sensor 1)
b. AL3 and AL4
chlorine, pH, temp, fault,
interval timer
temperature (sensor 1)
3. Deadband 0 to 9999 0
4. Interval timer settings
a. interval time
0.0 to 999.9 hr 24.0 hr
b. on time 0 to 999 sec 10 sec
c. recovery time 0 to 999 sec 60 sec