Emerson PN 51-FCL-1056 Marine Instruments User Manual


 
69
MODEL FCL-1056 SECTION 9.0
TROUBLESHOOTING
9.6.5 Calibration Was Successful, but Process pH Is Grossly Wrong and/or Noisy.
Grossly wrong or noisy readings suggest a ground loop (measurement system connected to earth ground at more
than one point), a floating system (no earth ground), or noise being brought into the analyzer by the sensor cable.
The problem arises from the process or installation. It is not a fault of the analyzer. The problem should disappear
once the sensor is taken out of the system. Check the following:
1. Is a ground loop present?
a. Verify that the system works properly in buffers. Be sure there is no direct electrical connection between the
buffer containers and the process liquid or piping.
b. Strip back the ends of a heavy gauge wire. Connect one end of the wire to the process piping or place it in
the process liquid. Place the other end of the wire in the container of buffer with the sensor. The wire makes
an electrical connection between the process and sensor.
c. If offsets and noise appear after making the connection, a ground loop exists.
2. Is the process grounded?
a. The measurement system needs one path to ground: through the process liquid and piping. Plastic piping,
fiberglass tanks, and ungrounded or poorly grounded vessels do not provide a path. A floating system can
pick up stray voltages from other electrical equipment.
b. Ground the piping or tank to a local earth ground.
c. If noise still persists, simple grounding is not the problem. Noise is probably being carried into the instrument
through the sensor wiring.
3. Simplify the sensor wiring.
a. Disconnect all sensor wires at the analyzer except, IN REFERENCE, IN pH, RTD IN and RTD RETURN.
See the wiring diagrams in Section 3.2.
b. Tape back the ends of the disconnected wires to keep them from making accidental connections with other
wires or terminals.
c. Connect a jumper wire between the RTD RETURN and RTD SENSE terminals (see wiring diagrams in
Section 3.2).
d. If noise and/or offsets disappear, the interference was coming into the analyzer through one of the sensor
wires. The system can be operated permanently with the simplified wiring.
4. Check for extra ground connections or induced noise.
a. To avoid induced noise in the sensor cable, keep the unit as far away as possible from power cables, relays,
and electric motors.
b. If ground loops persist, consult the factory. A visit from an experienced technician may be required to solve
the problem.
9.6.6 pH Readings Are Moderately Noisy and Tend to Wander.
pH readings that are moderately noisy (±0.1 pH) and tend to wander are probably caused by bubbles getting
trapped against the pH sensor. Although the overflow sampler is designed to allow bubbles to escape before they
reach the pH sensor and the sensor itself is designed so trapped air bubbles don’t interfere with the measurement,
bubbles may occasionally be a problem. Shaking the sensor will dislodge the bubbles. If bubbles remain a
problem, contact the factory.