Emerson PN 51-FCL-1056 Marine Instruments User Manual


 
35
MODEL FCL-1056 SECTION 5.0
PROGRAMMING THE ANALYZER
5.8.3 Procedure – Setting Up Diagnostics
1. Press MENU. The main menu screen appears. Move the cursor to
Program and press ENTER.
2. Scroll to the bottom of the screen and continuing scrolling until
Diagnostic Setup is highlighted. Press ENTER.
3. Diagnostics are available only for pH sensors. In the FCL-02 the pH sen-
sor is sensor 2. Press ENTER.
4. The screen summarizes the present diagnostic settings and limits. There
are nine items. To show items beyond the first four in the list, scroll to the
bottom of the list and continue scrolling.
To make a change, move the cursor to the desired line and press ENTER.
A screen will appear in which the present setting can be edited. It is
recommended that the settings be set to the values in the table:
5. To return to the main display, press MENU then EXIT.
Diagnostic Setup
Sensor 2
Sensor 2
Program
Measurement
Temperature
Security
Diagnostic Setup
S2 Diagnostic Setup
Ref Offset: 60mV
Diagnostic: On
Z Temp Corret’n: On
GI Fault 1500MΩ
GI Fault High:
Setting Default
Ref Offset 60 mV
Diagnostic On
Z Temp Correct’n On
Gl Fault High
1000 MΩ
Ref Fault High 20 KΩ