MODEL FCL-1056 SECTION 5.0
PROGRAMMING THE ANALYZER
5.4.3 Procedure – Configuring Alarms and Assigning Setpoints
1. Press MENU. The main menu screen appears. Move the cursor to
Program and press ENTER.
2. Choose Alarms.
3. Choose Configure/Setpoint.
4. Choose Alarm 1, Alarm 2, Alarm 3, or Alarm 4.
5. The screen summarizes the present configuration and setpoints. There
are nine items: Setpoint, Assign (S1 is sensor 1 and S2 is sensor 2),
Logic, Deadband, Interval time, On time, Recover time, and Hold
while active. The last four items describe the operation of the timer. Only
four items are shown at a time. To view the remaining items, scroll to the
bottom of the screen and continue scrolling.
To make a change, move the cursor to the desired line and press ENTER.
A screen will appear in which the present setting can be edited. Press
ENTER to store the setting.
For an explanation of terms, see sections 5.4.1 and 5.4.2.
6. To return to the main display, press MENU then EXIT.
Alarms
Configure/Setpoint
Simulate
Synch Timers: Yes
Configure/Setpoint
Configure/Setpoint
Alarm 1
Alarm 2
Alarm 3
Alarm 4
Alarm 1
Program
Outputs
Alarms
Measurement
Temperature
Alarms
Alarm 1 Settings
Setpoint: 0.000 ppm
Assign: S1 Measure
Logic: Low
Deadband: 0.000 ppm
Setpoint: