MODEL FCL-1056 SECTION 5.0
PROGRAMMING THE ANALYZER
5.7 CONFIGURING SECURITY SETTINGS
5.7.1 Purpose
This section describes how to set security codes. There are three levels of security.
a. A user can view the default display and diagnostic screens only.
b. A user has access to the calibration and hold menus only.
c. A user has access to all menus.
The security code is a three digit number. The table shows what happens when different security codes (XXX and
YYY) are assigned to Calibration/Hold and All. 000 means no security.
Calibration/Hold All
What happens…
000 XXX User enters XXX and has access to all menus.
XXX YYY
User enters XXX and has access to Calibration and Hold
menus only. User enters YYY and has access to all menus.
XXX 000 User needs no security code to have access to all menus.
000 000 User needs no security code to have access to all menus.
33
Temperature
Units: C
S1 Temp Comp: Auto
S2 Temp Comp: Auto
Units:
1. Press MENU. The main menu screen appears. Move the cursor to
Program and press ENTER.
2. Choose Temperature.
3. The screen summarizes the present sensor configuration. There will be
between three and five items. Units, S1 Temp Comp, and S2 Temp
Comp always appear. If manual temperature compensation was selected,
the manual temperature values entered for each sensor (S1 and S2
Manual) will also appear.
To make a change, move the cursor to the desired line and press
ENTER. A screen will appear in which the present setting can be
edited. To store a setting, press ENTER.
4. For an explanation of terms, see sections 5.6.2 and 5.6.3.
5. To return to the main display, press MENU then EXIT.
Program
Outputs
Alarms
Measurement
Temperature
Temperature
5.6.4 Procedure – Configuring Temperature Related Settings