Microplate Luminometer Maintenance
95
Insert the voltage selection card as shown in Figure 8-4,
pushing it into the instrument in the direction of the arrows,
with the printed side facing the power cord connector.
Adjust the indication pin in horizontal position.
Insert fuse holder again properly so that the indication pin fits
into the hole of the fuse holder with the proper operating volt-
age. It must work smoothly, otherwise the operating voltage is
not set correctly.
Function check:
Attach power cable.
Turn instrument on.
Check if green LED is on.
If the green LED is not on, please contact our Service Depart-
ment.
Use only UL-approved fuses. Otherwise, the operating safety
cannot be assured and the instrument will lose its UL ap-
proval.