Microplate Luminometer Installation and Commissioning
38
❏ The marked area is cleared again when you define a new
area with the mouse and click the
button again.
Running a measurement
❏ Click to start the measurement. The defined measure-
ment procedure begins: the individual wells are selected in
succession and measured using the predefined measure-
ment time. The results are displayed in the well matrix on the
screen. These are net results, provided a background meas-
urement has been carried out prior to the measurement
(display of the BG values in the field [Background]) and the
item [Subtract] has been activated.
❏ Click
to enter a comment for the measurement which will
be printed out in the report. This opens the [Measurement
Comment] dialog box and you can enter your text.
❏ Click
to save the measurement data and then enter a
file name.
Figure 3-17:
Defined well area