Microplate Luminometer Raw Data
56
When using two injectors:
Select both injectors. The following dialog box appears:
[Volume]: Enter
the injection volumes in µl (via keyboard or by
clicking on the arrow buttons).
[Injection 1 to injection 2 delay]: Delay time between the injec-
tions of injector 1 and 2. The hardware-dependent minimum
value of 2.05 seconds is defaulted and can be increased (t
P-I
).
[Injection 2 to measurement delay]: Enter the delay time be-
tween the injection of injector 2 and the start of the measure-
ment in seconds. 0 second is defaulted and can be increased (t
M-I
).
Example: Programming a measurement with 2 injections:
P = pre-injection
I=injection
M= measurement
t
P-
= delay between pre-injection and 2
nd
injection
t
M-I
= delay between 2
nd
injection and start of measurement
T
M
= duration of measurement
Figure 5-3:
[Timing] tab.
Selection o
f
injector 1 and 2
Injector 1 Injector 2