Microplate Luminometer Maintenance
93
8.3 Changing the Operating Voltage for
MONOLIGHT 3096
You need:
1 small flathead screwdriver
1 flat-nose pliers
2 fuses for the respective operating voltage
For operating voltage 230 V: Fuse: 160 mA slow-blow
For operating voltage 115/120 V: Fuse: 315 mA slow-blow
Use only UL-approved fuses!
The instrument is designed for operation at 230 V (Euro voltage)
or 115 V (US Voltage). The voltage setting must be changed only
when operating the instrument in a country with a different
operating voltage.
Power cord connector
0
100V
120V
230V
240V
I
Power switch Recess for screwdriver Fuse holder
Turn main switch to off ("0").
Unplug power cord from wall outlet.
Insert a screwdriver or a similar tool into the recess on the
right side of the fuse holder, push slightly and take the fuse
holder out.
Replace both fuses and carefully put the fuse holder aside.
Figure 8-2:
Power supply unit
on the instrument
rear panel