1 6-8
Process Flange –CF3M (Cast version of 316L SST)
Drain/Vent Valves –316L SST
Process Connections – 3/8-inch Swagelok compression fitting, 316L SST (1/4-18
NPT optional)
10/21/09
2 3-8 Change the first paragraph on the page to read as follows:
”Damping electronics are available as an option. Transmitters with standard
electronics can be retrofitted with the adjustable damping feature by changing out
both the amplifier board and the calibration board. Please reference Table 6-2
(Rosemount 1153 Series D Spare Parts List) for the applicable part numbers.”
4/13/12
3 6-11 Table 6-2,”Rosemount 1153 Series D Spare Parts List”, in all locations the
following part numbers are updated:
Amplifier Circuit Board, Output Code R:
”01154-0001-0005” is replaced by ”01154-0153-0001”
Amplifier Circuit Board with Damping, Output Code R:
”01154-0021-0004” is replaced by ”01154-0156-0001”
Amplifier Circuit Board, Output Code R, N0026:
”01154-0001-0006” is replace by ”01154-0153-0002”
Sensor Module, 316 SST
(5)
:
0-5/30 inH
2
O: ”01153-0320-0232” is replaced by ”01153-5320-0232”
”01153-0320-0132” is replaced by ”01153-5320-0132”
0-25/150 inH
2
O: ”01153-0320-0242” is replaced by ”01153-5320-0242”
”01153-0320-0342” is replaced by ”01153-5320-0342”
”01153-0320-0142” is replaced by ”01153-5320-0142”
0-125/750 inH
2
O: ”01153-0320-0252” is replaced by ”01153-5320-0252”
”01153-0320-0352” is replaced by ”01153-5320-0352”
”01153-0320-0152” is replaced by ”01153-5320-0152”
”01153-0320-0052” is replaced by ”01153-5320-0052”
0-17/100 psi: ”01153-0320-0262” is replaced by ”01153-5320-0262”
”01153-0320-0362” is replaced by ”01153-5320-0362”
”01153-0320-0162” is replaced by ”01153-5320-0162”
”01153-0320-0062” is replaced by ”01153-5320-0062”
0-50/300 psi: ”01153-0320-0272” is replaced by ”01153-5320-0272”
”01153-0320-0372” is replaced by ”01153-5320-0372”
”01153-0320-0172” is replaced by ”01153-5320-0172”
”01153-0320-0072” is replaced by ”01153-5320-0072”
0-170/1,000 psi: ”01153-0320-0282” is replaced by ”01153-5320-0282”
”01153-0320-0182” is replaced by ”01153-5320-0182”
”01153-0320-0082” is replaced by ”01153-5320-0082”
0-500/3,000 psi: ”01153-0320-0192” is replaced by ”01153-5320-0192”
0-1,000/4,000 psi: ”01153-0320-0102” is replaced by ”01153-5320-0102”
4/13/12
4 6-12 Table 6-2,”Rosemount 1153 Series D Spare Parts List”, the table footnotes are
updated to add note (5) which will read as follows:
(5) IMPORTANT NOTICE: To maintain a transmitter’s qualified configuration,
when purchasing or installing a new Sensor Module, Rosemount
Qualification report D2011019 must be carefully reviewed to verify that the
Sensor Module to be installed and the associated Amplifier Circuit Board in a
given transmitter is a qualified configuration. As detailed in the referenced
report, not all Sensor Module part numbers are considered qualified in conjunction
with certain Amplifier Circuit Boards.
4/13/12