Rosemount 1153 Series D
6-4
Rosemount 1153HD:
Maximum zero shift after 3,000 psi (20.68 MPa) overpressure:
Rosemount 1153GD and 1153AD:
Maximum zero shift after 2,000 psi (13.8 MPa)
overpressure:
Maximum zero shift after 4,500 psi (31.0 MPa) overpressure:
Maximum zero shift after 6,000 psi (41.34 MPa) overpressure:
Static Pressure Zero Effect
Rosemount 1153DD:
Per 1,000 psi (6.89 MPa):
Rosemount 1153HD:
Per 1,000 psi (6.89 MPa):
Static Pressure Span Effect
The effect is systematic and can be calibrated out for a particular pressure
before installation. Correction uncertainty equals ±0.5% of input reading/1,000
psi (6.89 MPa).
Power Supply Effect
Less than 0.005% of output span per volt
Range Code Overpressure Effect
4 ±1.0% of upper range limit
5 ±2.0% of upper range limit
6, 7 ±5.0% of upper range limit
Range Code Overpressure Effect
3, 4 ±0.25% of upper range limit
5–8 ±1.0% of upper range limit
Range Code Overpressure Effect
9 ±0.5% of upper range limit
Range Code Overpressure Effect
0 ±0.25% of upper range limit
Range Code Zero Effect
4, 5 ±0.2% of upper range limit
3, 6–8 ±0.5% of upper range limit
Range Code Zero Effect
4–7 ±0.66% of upper range limit